The Effect of Total Fields' Area and Dose Distribution in Step and Shoot IMRT on Gamma Passing Rate Using OCTAVIUS 4D-1500 Detector Phantom | ||
| Iranian Journal of Medical Physics | ||
| مقاله 1، دوره 18، شماره 4، مهر و آبان 2021، صفحه 226-231 اصل مقاله (1.26 M) | ||
| نوع مقاله: Original Paper | ||
| شناسه دیجیتال (DOI): 10.22038/ijmp.2020.44712.1690 | ||
| نویسندگان | ||
| Abdulrahman Mohammed Abdulbaqi1؛ Siham Sabah Abdullah2؛ Haydar Hamza Alabedi3؛ nabaa alazawy* 4؛ Mustafa Jabbar Al-Musawi5؛ Ahmed faris Heydar5 | ||
| 1Alamal National Hospital for Cancer Treatment, Baghdad, Iraq | ||
| 2Department of phsiology and medical physics, college of medicine, Alnahrain university, Iraq | ||
| 3Oncology Department, Baghdad University, Diwaniya, Iraq | ||
| 4Al-Mansour University College, Medical Instrumentation Engineering Department, Baghdad, Iraq | ||
| 5Physicist, Ministry of Health and Environment/Medical City, Baghdad Center for Radiotherapy and Nuclear Medicine, Baghdad, Iraq | ||
| چکیده | ||
| Introduction: Quality assurance is necessary for every IMRT plan.Octavius 4D-1500 detector phantom is one of the new phantoms for determining the treatment plan quality. This study aimed to examine the IMRT plans using the Octavius 4D-1500 to determine if it is a reliable, dependable, and durable. Material and Methods: IMRT QA conducted for 30 cases: HN and pelvis. The Monaco TPS used for treatment planning. The treatment plans were then applied to the Octavius 4D-1500 phantom (virtually and actually), the γ-index was calculated in VeriSoft program to evaluate the IMRT plans. Results: Significant differences were observed between the measured and calculated dose distributions for HN and pelvic plans, while the treatment sites did not affect the GP rate. The results of the global Gp were higher than the local GP, regardless of the study criteria. The HN plans showed a more significant difference than the pelvic plans. The HN plans, a strong significant correlation was found between the total fields’area and %GP in both global and local analyses, while in the pelvic plans, there was only a significant association with the local %GP. Conclusion: The measured dose distributions significantly different from calculated distributions. The relationship between the fields’ area and %GP was inverse. In the HN plans, a significant correlation found between the total fields’ area and %GP in both global and local, while only local %GP in the pelvic plans was significant correlation. Overall, the Octavius 4D-1500 detector phantom might be applicable for assessing the QA of IMRT plans. | ||
| کلیدواژهها | ||
| Gamma Passing Rate؛ Dose Distribution؛ Intensity Modulated Radiotherapy؛ Quality Assurance | ||