A comparison of 2D and 3D kidney absorbed dose measures in patients receiving 177Lu-DOTATATE | ||
Asia Oceania Journal of Nuclear Medicine and Biology | ||
مقاله 5، دوره 6، شماره 2، مهر 2018، صفحه 113-119 اصل مقاله (639.97 K) | ||
نوع مقاله: Original Article | ||
شناسه دیجیتال (DOI): 10.22038/aojnmb.2018.26105.1182 | ||
نویسندگان | ||
Kathy Willowson* 1؛ HyunJu Ryu2؛ Price Jackson3؛ Enid Eslick4؛ Anita Singh5؛ Dale Bailey6 | ||
1Institute of Medical Physics, The University of Sydney | ||
2Faculty of Health Sciences, University of Sydney, Lidcombe 2141, Australia | ||
3Department of Cancer Imaging, Peter MacCallum Cancer Centre, Melbourne 3000, VIC, Australia | ||
4Institute of Medical Physics, School of Physics, University of Sydney, 2006 NSW, Australia | ||
5Department of Nuclear Medicine, Royal North Shore Hospital, St Leonards 2065, NSW, Australia | ||
6Department of Nuclear Medicine, Royal North Shore Hospital, St Leonards 2065, NSW, Australia Faculty of Health Sciences, University of Sydney, Lidcombe 2141, Australia | ||
چکیده | ||
Objective(s): To investigate and compare quantitative accuracy of kidney absorbed dose measures made from both 2D and 3D imaging in patients receiving 177LuDOTATATE (Lutate) for treatment of neuroendocrine tumours (NETs). Methods: Patients receiving Lutate therapy underwent both whole body planar imaging and SPECT/CT imaging over the kidneys at time points 0.5, 4, 24, and 96-120 hours after injection. Planar data were corrected for attenuation using transmission data, and were converted to units of absolute activity via two methods, using either a calibration standard in the field of view or relative to pre-voiding image total counts. Hand drawn regions of interest were used to generate time activity curves and kidney absorbed dose estimates in OLINDA-EXM. Fully quantitative SPECT data were generated using CT-derived corrections for both scatter and attenuation, before correction for dead time and application of a camera specific sensitivity factor to convert data to units of absolute activity. Volumes of interest were defined for kidney using the co-registered x-ray CT, before time activity curves and absorbed dose measures were generated in OLINDA-EXM, both with and without corrections made to the model for patient specific kidney volumes. Quantitative SPECT data were also used to derive dose maps through dose kernel convolution (DKC), which was treated as the gold standard. Results: A total of 50 studies were analysed, corresponding to various cycles of treatment from 21 patients. Planar absorbed dose estimates were consistently higher than SPECT derived estimates by, on average, a factor of 3. Conclusion: Quantitative SPECT is considered the gold standard approach for organ specific dosimetry however often relies on in house software. As such planar methods for estimating absorbed dose are much more widely available, and in particular, are often the only source of reference in previously published data. For the case of Lutate dosimetry, planar measures may lead to a three-fold increase in measures of kidney absorbed dose. | ||
کلیدواژهها | ||
Lutate؛ Dosimetry؛ Kidney؛ NET | ||
مراجع | ||
10. Heikkonen J, Mäenpää H, Hippeläinen E, Reijonen V, Tenhunen M. Effect of calculation method on kidney dosimetry in 177Lu-octreotate treatment. Acta Oncol. 2016;55(9-10):1069-76.
11. Ljunberg M, Celler A, Konijnenberg MW, Eckerman K, Dewaraja YK, Sjogreen-Gleisner K. MIRD pamphlet No. 26: Joint EANM/MIRD guidelines for quantitative 177Lu SPECT applied for dosimetry of radiopharmaceutical therapy. J Nucl Med. 2016;57(1):151-62.
12. Bailey D, Hennessy T, Willowson K, Schembri G, Roach P, Snowdon G, et al. Quantitative biodistribution & kinetics of a new formulation of [Lu-177]-Octreotate. J Nucl Med. 2015;3:336.
13. Aslani A, Snowdon GM, Bailey DL, Schembri GP, Bailey EA, Pavlakis N, et al. Lutetium-177 DOTATATE Production with an Automated Radiopharmaceutical Synthesis System. Asia Ocean J Nucl Med Biol. 2015;3(2):107-15.
14. Siegel JA, Thomas SR, Stubbs JB, Stabin MG, Hays MT, Koral KF, et al. MIRD pamphlet no. 16: techniques for quantitative radiopharmaceutical biodistribution data acquisition and analysis for use in human radiation dose estimates. J Nucl Med. 1999;40(2):37S-61S.
15. Stabin MG, Sparks RB, Crowe E. OLINDA/EXM: the second-generation personal computer software for internal dose assessment in nuclear medicine. J Nucl Med. 2005;46(6):1023-7.
16. Willowson K, Bailey DL, Baldock C. Quantitative SPECT reconstruction using CT-derived corrections. Phys Med Biol. 2008;53(12):3099-112.
17. Jackson PA, Beauregard JM, Hofman MS, Kron T, Hogg A, Hicks RJ. An automated voxelized dosimetry tool for radionuclide therapy based on serial quantitative SPECT/CT imaging. Med Phys. 2013;40(11):112503. | ||
آمار تعداد مشاهده مقاله: 1,731 تعداد دریافت فایل اصل مقاله: 887 |